Feasibility of Monitoring Heart and Respiratory Rates Using Nonwearable Devices and Consistency of the Measured Parameters: Pilot Feasibility Study

Ikuta K, Aishima M, Noguchi-Watanabe M, Fukui S

Feasibility of Monitoring Heart and Respiratory Rates Using Nonwearable Devices and Consistency of the Measured Parameters: Pilot Feasibility Study

JMIR Hum Factors 2024;11:e56547

DOI: 10.2196/56547

PMID: 39378444

PMCID: 11479369

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Feasibility of monitoring heart and respiratory rates using non-wearable devices and consistency of the measured parameters: A pilot feasibility study

  • Kasumi Ikuta; 
  • Miya Aishima; 
  • Maiko Noguchi-Watanabe; 
  • Sakiko Fukui

ABSTRACT

Background:

As Japan is the world’s fastest-aging society with a declining population, it is challenging to secure human resources for care providers. Therefore, the Japanese government is promoting digital transformation (DX) and the use of nursing care equipment, including non-wearable devices that monitor heart and respiratory rates. However, the feasibility of monitoring heart and respiratory rates with non-wearable devices and the consistency of the rates measured have not been reported.

Objective:

In this study, we focus on a sheet-type non-wearable device (Safety Sheep Sensor) that has been introduced in many nursing homes. We evaluated the feasibility of monitoring heart rate (HR) and respiratory rate (RR) continuously using non-wearable devices and the consistency of the HR and RR measured.

Methods:

A sheet-type non-wearable device (Safety Sheep Sensor) that measured HR and RR every minute through body vibrations was placed under the mattress of each participant. The participants in Study 1 were healthy individuals aged 20–60 years (N=21), while those in Study 2 were older individuals living in multi-dwelling houses and required nursing care (N=20). The HR was measured using standard methods by the nurse and using the wearable device (WD; Silmee Bar type Lite sensor), and RR was measured by the nurse. The primary outcome was the mean difference in HR and RR between non-wearable devices and standard methods.

Results:

The mean difference in HR was -0.32 (standard deviation (SD): 3.12) in Study 1 and 0.04 (SD: 3.98) in Study 2; both differences were within the predefined accepted discrepancies (< 5 beats). The mean difference in RR was -0.98 (SD: 3.01) in Study 1 and -0.49 (SD: 2.40) in Study 2; both differences were within the predefined accepted discrepancies (3 breaths).

Conclusions:

HR and RR measurements obtained using the non-wearable devices and the standard method were similar. Continuous monitoring of vital signs using non-wearable devices can aid in the early detection of abnormal conditions in older people.


 Citation

Please cite as:

Ikuta K, Aishima M, Noguchi-Watanabe M, Fukui S

Feasibility of Monitoring Heart and Respiratory Rates Using Nonwearable Devices and Consistency of the Measured Parameters: Pilot Feasibility Study

JMIR Hum Factors 2024;11:e56547

DOI: 10.2196/56547

PMID: 39378444

PMCID: 11479369

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.