Accepted for/Published in: JMIR mHealth and uHealth

Date Submitted: Oct 13, 2023
Date Accepted: Jun 25, 2024

The final, peer-reviewed published version of this preprint can be found here:

Reliable Contactless Monitoring of Heart Rate, Breathing Rate, and Breathing Disturbance During Sleep in Aging: Digital Health Technology Evaluation Study

Guruswamy Ravindran KK, della Monica C, Atzori G, Lambert D, Hassanin H, Revell V, Dijk DJ

Reliable Contactless Monitoring of Heart Rate, Breathing Rate, and Breathing Disturbance During Sleep in Aging: Digital Health Technology Evaluation Study

JMIR Mhealth Uhealth 2024;12:e53643

DOI: 10.2196/53643

PMID: 39190477

PMCID: 11387924

Reliable Contactless Monitoring of Heart Rate, Breathing Rate and Breathing Disturbance During Sleep in Aging: A Digital Health Technology Evaluation Study

  • Kiran Kumar Guruswamy Ravindran; 
  • Ciro della Monica; 
  • Giuseppe Atzori; 
  • Damion Lambert; 
  • Hana Hassanin; 
  • Victoria Revell; 
  • Derk-Jan Dijk

ABSTRACT

Background:

Longitudinal monitoring of vital signs provides a method for identifying changes to general health in an individual and particularly so in older adults. The nocturnal sleep period provides a convenient opportunity to assess vital signs. Contactless technologies that can be embedded into the bedroom environment are unintrusive and burdenless and have the potential to enable seamless monitoring of vital signs. To realise this potential, these technologies need to be evaluated against gold standard measures and in relevant populations.

Objective:

We evaluated the accuracy of heart rate and breathing rate measurements of three contactless technologies (two under-mattress trackers: Withings sleep analyser (WSA) and Emfit QS (Emfit) and a bedside radar: Somnofy) in a sleep laboratory environment and assessed their potential to capture vital signs in a real-world setting.

Methods:

Data were collected in 35 community dwelling older adults aged between 65 and 83 years (mean ± SD: 70.8 ± 4.9; 21 men) during a one-night clinical polysomnography (PSG) in a sleep laboratory, preceded by 7 to 14 days of data collection at-home. Several of the participants had health conditions including type-2 diabetes, hypertension, obesity, and arthritis and ≈49% (n = 17) had moderate to severe sleep apnea while ≈29% (n = 10) had periodic leg movement disorder. The under-mattress trackers provided estimates of both heart rate and breathing rate while the bedside radar provided only breathing rate. The accuracy of the heart rate and breathing rate estimated by the devices was compared to PSG electrocardiogram (ECG) derived heart rate (beats per minute, bpm) and respiratory inductance plethysmography thorax (RIP thorax) derived breathing rate (cycles per minute, cpm). We also evaluated breathing disturbance indices of snoring and the apnea-hypopnea index (AHI) available from the WSA.

Results:

All three contactless technologies provided acceptable accuracy in estimating heart rate [mean absolute error (MAE) < 2.12 bpm and mean absolute percentage error (MAPE) < 5%] and breathing rate (MAE ≤ 1.6 cpm and MAPE < 12%) at 1-minute resolution. All three contactless technologies were able to capture changes in heart rate and breathing rate across the sleep period. The WSA snoring and breathing disturbance estimates were also accurate compared to PSG estimates (R-squared: WSA Snore: 0.76, p < 0.001; WSA AHI: 0.59, p < 0.001).

Conclusions:

Contactless technologies offer an unintrusive alternative to conventional wearable technologies for reliable monitoring of heart rate, breathing rate, and sleep apnea in community dwelling older adults at scale. They enable assessment of night-to-night variation in these vital signs, which may allow the identification of acute changes in health, and longitudinal monitoring which may provide insight into health trajectories.


 Citation

Please cite as:

Guruswamy Ravindran KK, della Monica C, Atzori G, Lambert D, Hassanin H, Revell V, Dijk DJ

Reliable Contactless Monitoring of Heart Rate, Breathing Rate, and Breathing Disturbance During Sleep in Aging: Digital Health Technology Evaluation Study

JMIR Mhealth Uhealth 2024;12:e53643

DOI: 10.2196/53643

PMID: 39190477

PMCID: 11387924

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.