ChroniSense National Early Warning Score Study: Comparison Study of a Wearable Wrist Device to Measure Vital Signs in Patients Who Are Hospitalized

Van Velthoven MH, Oke J, Kardos A

ChroniSense National Early Warning Score Study: Comparison Study of a Wearable Wrist Device to Measure Vital Signs in Patients Who Are Hospitalized

J Med Internet Res 2023;25:e40226

DOI: 10.2196/40226

PMID: 36745491

PMCID: 9941897

ChroniSense National Early Warning Score Study: Comparison Study of a Wearable Wrist Device to Measure Vital Signs in Patients Who Are Hospitalized

  • Michelle Helena Van Velthoven; 
  • Jason Oke; 
  • Attila Kardos

Background:

Wearable devices could be used to continuously monitor vital signs in patients who are hospitalized, but they require validation.

Objective:

This study aimed to evaluate the clinical validity of the prototype of a semiautomated wearable wrist device (ChroniSense Polso) to measure vital signs and provide National Early Warning Scores (NEWSs).

Methods:

Vital signs and NEWSs measured using the wearable device were compared with standard, nurse-lead manual measurements. We enrolled adult patients (aged ≥18 years) who required vital sign measurements at least every 6 hours in a UK teaching district general hospital. Wearable device measurements were not used for clinical decision-making. The primary outcome was the agreement on the individual National Early Warning parameter scores and vital sign measurements: respiratory rate, oxygen saturation, body temperature, systolic blood pressure, and heart rate. Secondary outcomes were the agreement on the total NEWS, incidence of adverse events, and user acceptance. To compare the wearable device measurements with the standard measurements, we analyzed vital sign measurements by limits of agreement (Bland-Altman analysis) and conducted κ agreement analyses for NEWSs. A user experience survey was conducted with questions about comfort of the wrist device, safety, preference, and use.

Results:

We included 132 participants in the study, with a mean age of 62 (SD 15.81) years; most of them were men (102/132, 77.3%). The highest weighted κ values were found for heart rate (0.69, 95% CI 0.57-0.81 for all 385 measurements) and systolic blood pressure (0.39, 95% CI 0.30-0.47 for all 339 measurements). Weighted κ values were low for respiration rate (0.03, 95% CI −0.001 to 0.05 for all 445 measurements), temperature (0, 95% CI 0-0 for all 231 measurements), and oxygen saturation (−0.11, 95% CI −0.20 to −0.02 for all 187 measurements). Weighted κ using Cicchetti-Allison weights showed κ of 0.20 (95% CI 0.03-0.38) when using all 56 total NEWSs. The user acceptance survey found that approximately half (45/91, 49%) of the participants found it comfortable to wear the device and liked its appearance. Most (85/92, 92%) of them said that they would wear the device during their next hospital visit, and many (74/92, 80%) said that they would recommend it to others.

Conclusions:

This study shows the promising use of a prototype wearable device to measure vital signs in a hospital setting. Agreement between the standard measurements and wearable device measurements was acceptable for systolic blood pressure and heart rate, but needed to be improved for respiration rate, temperature, and oxygen saturation. Future studies need to improve the clinical validity of this wearable device. Large studies are required to assess clinical outcomes and cost-effectiveness of wearable devices for vital sign measurement.

International Registered Report:

RR2-10.1136/bmjopen-2018-028219


 Citation

Please cite as:

Van Velthoven MH, Oke J, Kardos A

ChroniSense National Early Warning Score Study: Comparison Study of a Wearable Wrist Device to Measure Vital Signs in Patients Who Are Hospitalized

J Med Internet Res 2023;25:e40226

DOI: 10.2196/40226

PMID: 36745491

PMCID: 9941897

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.