A Remote Patient-Monitoring System for Intensive Care Medicine: Mixed Methods Human-Centered Design and Usability Evaluation

Poncette AS, Mosch L, Stablo L, Spies C, Schieler M, Weber-Carstens S, Feufel MA, Balzer F

A Remote Patient-Monitoring System for Intensive Care Medicine: Mixed Methods Human-Centered Design and Usability Evaluation

JMIR Hum Factors 2022;9(1):e30655

DOI: 10.2196/30655

PMID: 35275071

PMCID: 8957007

Human-Centered Design with Usability Evaluation of a Remote Patient Monitoring System for Intensive Care Medicine: A Mixed-Methods Study

  • Akira-Sebastian Poncette; 
  • Lina Mosch; 
  • Lars Stablo; 
  • Claudia Spies; 
  • Monique Schieler; 
  • Steffen Weber-Carstens; 
  • Markus A. Feufel; 
  • Felix Balzer

ABSTRACT

Background:

Continuous monitoring of vital signs is critical for ensuring patient safety in intensive care units (ICUs) and is becoming increasingly relevant in general wards. The effectiveness of health information technologies such as patient monitoring systems is highly determined by usability, the lack of which can ultimately compromise patient safety. Usability problems can be identified and prevented by involving users (ie, clinicians).

Objective:

In this study, we applied a human-centered design (HCD) approach to evaluate the usability of a remote patient monitoring system user interface (UI) in the ICU context, and to conceptualize and evaluate design changes.

Methods:

Following IRB approval (EA1/031/18), a formative evaluation of the monitoring UI was performed. Simulated use tests with think-aloud protocols were conducted with ICU staff (n = 5) and the resulting qualitative data was analyzed using a deductive analytic approach. Based on the identified usability problems, we conceptualized informed design changes and applied them to develop an improved prototype of the monitoring UI. Comparing the UIs, we evaluated perceived usability with the system usability scale (SUS), performance efficiency with the normative path deviation (NPD), and effectiveness by measuring the task completion rate. Measures were tested for statistical significance using a two-sample t test, Poisson regression with a generalized linear mixed-effects model, and the N-1 chi-square test. P values <.05 were considered significant.

Results:

We found 37 individual usability problems specific to the monitoring UI, which could be assigned to 6 subcodes: Usefulness of the system, response time, responsiveness, meaning of labels, function of UI elements, and navigation. Among user ideas and requirements for the UI were high usability, customizability, and the provision of audible alarm notifications. Changes in graphics and design were proposed to allow better navigation, information retrieval, and spatial orientation. The UI was revised by creating a prototype with a more responsive design and changes regarding labeling and UI elements. Statistical analysis showed that perceived usability improved significantly (SUS Design A: mean 68.5, SD 11.26, n = 5; Design B: mean 89, SD 4.87, n = 5; P = .003), as did performance efficiency (NPD Design A: mean 8.8, SD 5. 26, n = 5; Design B: mean 3.2, SD 3.03, n = 5; P = .001), and effectiveness (Design A: 18 trials, failed 7 times, passed 11 times; Design B: 20 trials, failed 0 times, passed 20 times; P = .002).

Conclusions:

Usability testing with think-aloud protocols led to a patient monitoring UI with a significantly improved usability, performance and effectiveness. In the ICU work environment, difficult-to-use technology may result in detrimental outcomes for staff and patients. Technical devices should be designed to support efficient and effective work processes. Our results suggest that this can be achieved by applying basic HCD methods and principles. Clinical Trial: ClinicalTrials.gov NCT03514173


 Citation

Please cite as:

Poncette AS, Mosch L, Stablo L, Spies C, Schieler M, Weber-Carstens S, Feufel MA, Balzer F

A Remote Patient-Monitoring System for Intensive Care Medicine: Mixed Methods Human-Centered Design and Usability Evaluation

JMIR Hum Factors 2022;9(1):e30655

DOI: 10.2196/30655

PMID: 35275071

PMCID: 8957007

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.