Maintenance Notice

Due to necessary scheduled maintenance, the JMIR Publications website will be unavailable from Wednesday, July 01, 2020 at 8:00 PM to 10:00 PM EST. We apologize in advance for any inconvenience this may cause you.

Who will be affected?

Accepted for/Published in: Journal of Medical Internet Research

Date Submitted: Mar 21, 2021
Date Accepted: Nov 21, 2021

The final, peer-reviewed published version of this preprint can be found here:

The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study

Santos M, Vollam S, Pimentel MA, Areia C, Young L, Roman C, Ede J, Piper P, King E, Harford M, Shah A, Gustafson O, Tarassenko L, Watkinson P

The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study

J Med Internet Res 2022;24(2):e28890

DOI: 10.2196/28890

PMID: 35166690

PMCID: 8889481

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Wearable pulse oximeters in the prompt detection of hypoxaemia and during movement: a diagnostic accuracy study

  • Mauro Santos; 
  • Sarah Vollam; 
  • Marco A.F. Pimentel; 
  • Carlos Areia; 
  • Louise Young; 
  • Cristian Roman; 
  • Jody Ede; 
  • Philippa Piper; 
  • Elizabeth King; 
  • Mirae Harford; 
  • Akshay Shah; 
  • Owen Gustafson; 
  • Lionel Tarassenko; 
  • Peter Watkinson

ABSTRACT

Background:

Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms, however, their reliability is unclear to use in the hospital setting.

Objective:

We report the performance of wearable pulse oximeters in a simulated clinical setting when challenged by motion and low levels of arterial blood oxygen (SaO2).

Methods:

The performance of one wrist-worn (Wavelet) and three finger-worn (CheckMeTM O2+, AP-20 and WristOx2® 3150) wearable, wireless transmission-mode, pulse oximeters was evaluated. Seven motion tasks were performed: At rest, Sit-to-Stand, Tapping, Rubbing, Drinking, Turning Pages, and Using a Tablet. Hypoxia exposure followed, in which inspired gases were adjusted to achieve decreasing SaO2 levels at 100%, 95%, 90%, 87%, 85%, 83% and 80%. Peripheral oxygen saturation (SpO2) estimates were compared with simultaneous SaO2 samples to calculate the root mean squared error (RMSE). Area under the receiver-operating characteristic curve was used to analyse the detection of hypoxaemia, SaO2 < 90%.

Results:

SpO2 estimates matching 215 SaO2 samples in both study phases, from 33 participants, were analysed. Tapping, rubbing, turning pages and using a tablet degraded SpO2 estimation (RMSE > 4% for a least one device). All finger-worn pulse oximeters detected hypoxaemia, with an overall sensitivity ≥ 0.87 and specificity ≥ 0.80, comparable to that of the Philips MX450.

Conclusions:

The SpO2 accuracy of wearable finger-worn pulse oximeters was within that required by the International Organization for Standardization guidelines. Performance was degraded by motion, but all were capable of detecting hypoxaemia. Our findings support the use of wearable, wireless transmission-mode, pulse oximeters to detect the onset of clinical deterioration in hospital settings. Clinical Trial: ISRCTN61535692; http://www.isrctn.com/ISRCTN61535692


 Citation

Please cite as:

Santos M, Vollam S, Pimentel MA, Areia C, Young L, Roman C, Ede J, Piper P, King E, Harford M, Shah A, Gustafson O, Tarassenko L, Watkinson P

The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study

J Med Internet Res 2022;24(2):e28890

DOI: 10.2196/28890

PMID: 35166690

PMCID: 8889481

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.