Previously submitted to: Journal of Medical Internet Research (no longer under consideration since Feb 17, 2021)
Date Submitted: Feb 6, 2021
Open Peer Review Period: Feb 5, 2021 - Feb 17, 2021
(closed for review but you can still tweet)
NOTE: This is an unreviewed Preprint
Warning: This is a unreviewed preprint (What is a preprint?). Readers are warned that the document has not been peer-reviewed by expert/patient reviewers or an academic editor, may contain misleading claims, and is likely to undergo changes before final publication, if accepted, or may have been rejected/withdrawn (a note "no longer under consideration" will appear above).
Peer review me: Readers with interest and expertise are encouraged to sign up as peer-reviewer, if the paper is within an open peer-review period (in this case, a "Peer Review Me" button to sign up as reviewer is displayed above). All preprints currently open for review are listed here. Outside of the formal open peer-review period we encourage you to tweet about the preprint.
Citation: Please cite this preprint only for review purposes or for grant applications and CVs (if you are the author).
Final version: If our system detects a final peer-reviewed "version of record" (VoR) published in any journal, a link to that VoR will appear below. Readers are then encourage to cite the VoR instead of this preprint.
Settings: If you are the author, you can login and change the preprint display settings, but the preprint URL/DOI is supposed to be stable and citable, so it should not be removed once posted.
Submit: To post your own preprint, simply submit to any JMIR journal, and choose the appropriate settings to expose your submitted version as preprint.
Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.
Radiotherapy dose calculations in high-Z materials: comprehensive comparison between experiment, Monte Carlo, and conventional planning algorithms
ABSTRACT
Background:
Despite spinal metastases accounting for 10% to 30% of new tumors diagnoses annually, and radiation therapy is a standard treatment technique, the studies discussing the effects of small-size spinal prostheses on spinal radiation therapy are limited.
Objective:
To compare the accuracies of the AAA and AcurosXB dose calculation algorithms and to predict the change in the down-stream and lateral dose deposition of high energy photons in the presence of material with densities higher that commonly found in the body.
Methods:
Metal rods of titanium (d =4.5g/m2), stainless steel (d=8g/cm2) and tungsten (d=19.25 g/cm2) were positioned in a phantom. Film was position behind and laterally to the rods to measure the dose distribution for a 6 MV, 18 MV and 10 FFF photon beams. A DOSXYZnrc Monte Carlo simulation of the experimental setup was performed The AAA and AcurosXB dose calculation algorithms were used to predict the dose distributions. The dose from film and DOSXYZnrc were compared with the dose predicted by AAA and AcurosXB.
Results:
AAA overestimated the dose behind the rods by 15-25% and underestimated the dose laterally to the rods by 5-15% depending on the range of materials and energies investigated. AcurosXB overestimated the dose behind the rods by 1-18% and underestimated the dose laterally to the rods by up to 5% depending on the range of material and energies investigated.
Conclusions:
AAA cannot deliver clinically acceptable dose calculation results at a distance less than 10 mm from metals, for a single field treatment. Acuros XB is able to handle metals of low atomic numbers (Z ≤ 26), but not tungsten (Z = 74). This can be due to the restriction of the CT-density table in EclipseTM TPS, which has an upper HU limit of 10501.
Citation
Request queued. Please wait while the file is being generated. It may take some time.
Copyright
© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.