Accepted for/Published in: JMIR mHealth and uHealth

Date Submitted: Jan 26, 2021
Date Accepted: May 10, 2021

The final, peer-reviewed published version of this preprint can be found here:

Enabling Wearable Pulse Transit Time-Based Blood Pressure Estimation for Medically Underserved Areas and Health Equity: Comprehensive Evaluation Study

Ganti VG, Carek AM, Jung H, Srivatsa AV, Cherry D, Johnson LN, Inan OT

Enabling Wearable Pulse Transit Time-Based Blood Pressure Estimation for Medically Underserved Areas and Health Equity: Comprehensive Evaluation Study

JMIR Mhealth Uhealth 2021;9(8):e27466

DOI: 10.2196/27466

PMID: 34338646

PMCID: 8369375

Enabling Wearable Pulse Transit Time based Blood Pressure Estimation for Medically Underserved Areas and Health Equity: A Comprehensive Evaluation Study

  • Venu G Ganti; 
  • Andrew M Carek; 
  • Hewon Jung; 
  • Adith V Srivatsa; 
  • Deborah Cherry; 
  • Levather Neicey Johnson; 
  • Omer T Inan

ABSTRACT

Background:

Noninvasive approaches to monitor blood pressure (BP), in light of their convenience and accuracy, have paved the way towards remote screening and management of hypertension. However, existing noninvasive methodologies—which operate on mechanical, electrical, and optical sensing modalities—have not been thoroughly evaluated in demographically and racially diverse populations. Thus, the potential accuracy of these technologies in the very populations where they could have the most impact has not been sufficiently addressed. This presents challenges towards clinical translation, due to concerns of perpetuating existing health disparities.

Objective:

Here, we present findings on the feasibility of a cuff-less, wrist-worn, pulse transit time (PTT) based device for monitoring BP in a diverse population.

Methods:

The diverse population was recruited through a collaborative effort with a non-profit working with medically underserved areas (MUA) in Georgia. We used our custom, multi-modal, wrist-worn device to measure the PTT through photoplethysmography—as the distal timing reference—and seismocardiography—as the proximal timing reference. In addition, we created a novel data-driven beat selection algorithm to reduce noise and improve the robustness of the method. We compared the wearable PTT measurements to those from a finger-cuff continuous BP device over the course of several perturbations used to modulate BP.

Results:

Our PTT-based wrist-worn device accurately monitors BP in a diverse population (N = 44 subjects) with a mean-absolute-difference (MAD) of 2.90 mmHg after calibration. We further show the ability of our device to capture commonly observed demographic differences in underlying arterial stiffness.

Conclusions:

Accurate BP estimation using a convenient wearable device can empower users and facilitate remote blood pressure monitoring in MUAs, thus providing widespread hypertension screening and management towards health equity.


 Citation

Please cite as:

Ganti VG, Carek AM, Jung H, Srivatsa AV, Cherry D, Johnson LN, Inan OT

Enabling Wearable Pulse Transit Time-Based Blood Pressure Estimation for Medically Underserved Areas and Health Equity: Comprehensive Evaluation Study

JMIR Mhealth Uhealth 2021;9(8):e27466

DOI: 10.2196/27466

PMID: 34338646

PMCID: 8369375

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.