Maintenance Notice

Due to necessary scheduled maintenance, the JMIR Publications website will be unavailable from Wednesday, July 01, 2020 at 8:00 PM to 10:00 PM EST. We apologize in advance for any inconvenience this may cause you.

Who will be affected?

Accepted for/Published in: JMIR mHealth and uHealth

Date Submitted: Oct 9, 2020
Date Accepted: Nov 12, 2021
Date Submitted to PubMed: Jan 4, 2022

The final, peer-reviewed published version of this preprint can be found here:

Continuous Noninvasive Remote Automated Blood Pressure Monitoring With Novel Wearable Technology: A Preliminary Validation Study

McGillion M, Dvirnik N, Yang S, Belley-Côté E, Lamy A, Whitlock R, Marcucci M, Borges F, Duceppe E, Ouellette C, Bird M, Carroll SL, Conen D, Tarride JE, Harsha P, Scott T, Good A, Gregus K, Sanchez K, Benoit P, Owen J, Harvey V, Peter E, Petch J, Vincent J, Graham M, Devereaux P

Continuous Noninvasive Remote Automated Blood Pressure Monitoring With Novel Wearable Technology: A Preliminary Validation Study

JMIR Mhealth Uhealth 2022;10(2):e24916

DOI: 10.2196/24916

PMID: 34876396

PMCID: 8922156

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Continuous non-invasive remote automated blood pressure monitoring with novel wearable technology: A Validation Study

  • Michael McGillion; 
  • Nazari Dvirnik; 
  • Stephen Yang; 
  • Emilie Belley-Côté; 
  • Andre Lamy; 
  • Richard Whitlock; 
  • Maura Marcucci; 
  • Flavia Borges; 
  • Emmanuelle Duceppe; 
  • Carley Ouellette; 
  • Marissa Bird; 
  • Sandra L Carroll; 
  • David Conen; 
  • Jean-Eric Tarride; 
  • Prathiba Harsha; 
  • Ted Scott; 
  • Amber Good; 
  • Krysten Gregus; 
  • Karla Sanchez; 
  • Pamela Benoit; 
  • Julian Owen; 
  • Valerie Harvey; 
  • Elizabeth Peter; 
  • Jeremy Petch; 
  • Jessica Vincent; 
  • Michelle Graham; 
  • PJ Devereaux

ABSTRACT

Background:

Background:

Continuous hemodynamic monitoring is the standard of care for patients intraoperatively, but vital signs monitoring is performed only periodically on post-surgical wards, and patients are routinely discharged home with no surveillance. Wearable continuous monitoring biosensor technologies have the potential to transform postoperative care with early detection of impending clinical deterioration.

Objective:

Objective:

Our aim was to validate the accuracy of the Cloud DX Vitaliti™ Continuous Vital Signs Monitor (CVSM) continuous non-invasive blood pressure measurements in post-surgical patients. Usability of the Vitaliti™ CVSM was also examined.

Methods:

Methods:

Included patients were recovering from surgery in a cardiac intensive care unit. Validation procedures were according to AAMI - ISO 81060-2 2013 standards for Wearable, Cuffless Blood Pressure Measuring Devices. In static (seated in bed) and supine positons, three 30-second cNIBP measurements were taken for each patient with the Vitaliti™ CVSM and an invasive arterial catheter. The errors of these determinations were calculated. Participants were interviewed about device acceptability

Results:

Results:

Data for 21 patients were included in the validation analysis. The overall mean and SD of the errors of determination for the static position were -0.784 mmHg (SD 4.594) for systolic blood pressure and 0.477 mmHg (SD 1.668) for diastolic blood pressure. Errors of determination were slightly higher for the supine position at 3.533 mmHg (SD 6.335) for systolic blood pressure and 3.050 mmHg (SD 3.619) for diastolic blood pressure. The majority rated the Vitaliti™ CVSM as comfortable.

Conclusions:

Conclusion: The Vitaliti™ CVSM was compliant with AAMI-ISO 81060-2:2013 standards and well-received by patients. Clinical Trial: Trial Registration: ClinicalTrials.gov (NCT03493867)


 Citation

Please cite as:

McGillion M, Dvirnik N, Yang S, Belley-Côté E, Lamy A, Whitlock R, Marcucci M, Borges F, Duceppe E, Ouellette C, Bird M, Carroll SL, Conen D, Tarride JE, Harsha P, Scott T, Good A, Gregus K, Sanchez K, Benoit P, Owen J, Harvey V, Peter E, Petch J, Vincent J, Graham M, Devereaux P

Continuous Noninvasive Remote Automated Blood Pressure Monitoring With Novel Wearable Technology: A Preliminary Validation Study

JMIR Mhealth Uhealth 2022;10(2):e24916

DOI: 10.2196/24916

PMID: 34876396

PMCID: 8922156

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.