Accepted for/Published in: JMIR Medical Informatics

Date Submitted: May 1, 2020
Date Accepted: Jul 3, 2020

The final, peer-reviewed published version of this preprint can be found here:

Integrating and Evaluating the Data Quality and Utility of Smart Pump Information in Detecting Medication Administration Errors: Evaluation Study

Ni Y, Lingren T, Huth H, Timmons K, Melton K, Kirkendall E

Integrating and Evaluating the Data Quality and Utility of Smart Pump Information in Detecting Medication Administration Errors: Evaluation Study

JMIR Med Inform 2020;8(9):e19774

DOI: 10.2196/19774

PMID: 32876578

PMCID: 7495258

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Integrating and Evaluating the Usability and Utility of Smart Pump Information in Detecting Medication Administration Errors

  • Yizhao Ni; 
  • Todd Lingren; 
  • Hannah Huth; 
  • Kristen Timmons; 
  • Krisin Melton; 
  • Eric Kirkendall

ABSTRACT

Background:

At present, electronic health records (EHRs) are the central focus of clinical informatics given their role as the primary source of clinical data. Despite their granularity, the EHR data heavily rely on manual input and are prone to human errors. Many other sources of data exist in the clinical setting, including digital medical devices such as smart infusion pumps. When incorpo-rated with prescribing data from the EHRs, smart pump records (SPR) are capable of shedding light on actions taking place during the medication use process. Nevertheless, harmonizing the two sources is hindered by multiple technical challenges, and the usability and utility of SPRs have not been realized.

Objective:

In this study we incorporated SPRs with EHR data to evaluate their usability and utility in detecting medication administration errors. Our overarching hypothesis was that SPRs would contribute unique information in the medication use process, enabling more comprehensive detection of discrepancies and potential errors in medication administration.

Methods:

We evaluated the medication use process of nine high-risk medications for patients admitted to the neonatal intensive care unit during a one-year period. An automated algorithm was developed to align SPRs with their medication orders in the EHRs using patient ID, medication name and timestamp. The aligned data were manually reviewed by a clinical research coordinator and two pediatric physicians to identify discrepancies in medication administration. The usability of SPRs was assessed with the proportion of usable data that were linked to valid EHR orders. To evaluate their utility, we compared the frequency and severity of discrepancies captured by the SPR and EHR data respectively. A novel concordance assessment was also developed to under-stand the detection power and capabilities of SPR and EHR data.

Results:

Approximately 70% of the SPRs contained valid patient IDs and medication names, making them usable for data integration. After combing the two sources, the investigative team re-viewed 2307 medication orders with 10575 medication administration records (MARs) and 23397 SPRs. A total of 321 MAR and 682 SPR discrepancies were identified, with vasopressors showing the highest discrepancy rates, followed by narcotics and total parenteral nutrition. Compared to EHR MARs, substantial dosing discrepancies were more detectable using the SPRs. The concordance analysis showed little overlap between MAR and SPR discrepancies, with most discrepancies captured by the SPR data.

Conclusions:

We integrated smart infusion pump information with EHR data to make visible the most error-prone phases of the medication lifecycle. The findings suggested that SPRs could be a more re-liable data source for medication error detection. Ultimately, it is imperative to integrate SPR information with EHR data to fully detect and mitigate medication administration errors in the clinical setting.


 Citation

Please cite as:

Ni Y, Lingren T, Huth H, Timmons K, Melton K, Kirkendall E

Integrating and Evaluating the Data Quality and Utility of Smart Pump Information in Detecting Medication Administration Errors: Evaluation Study

JMIR Med Inform 2020;8(9):e19774

DOI: 10.2196/19774

PMID: 32876578

PMCID: 7495258

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.