Accuracy of the Multisensory Wristwatch Polar Vantage's Estimation of Energy Expenditure in Various Activities: Instrument Validation Study

Gilgen-Ammann R, Schweizer T, Wyss T

Accuracy of the Multisensory Wristwatch Polar Vantage's Estimation of Energy Expenditure in Various Activities: Instrument Validation Study

JMIR Mhealth Uhealth 2019;7(10):e14534

DOI: 10.2196/14534

PMID: 31579020

PMCID: 6777286

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Accuracy of the Multisensory Wristwatch Polar Vantage's Estimation of Energy Expenditure in Various Activities: Instrument Validation Study

  • Rahel Gilgen-Ammann; 
  • Theresa Schweizer; 
  • Thomas Wyss

Background:

Sport watches and fitness trackers provide a feasible way of obtaining energy expenditure (EE) estimations in daily life as well as during exercise. However, today’s popular wrist-worn technologies show only poor-to-moderate EE accuracy. Recently, the invention of optical heart rate measurement and the further development of accelerometers in wrist units have opened up the possibility of measuring EE.

Objective:

This study aimed to validate the new multisensory wristwatch Polar Vantage and its EE estimation in healthy individuals during low-to-high-intensity activities against indirect calorimetry.

Methods:

Overall, 30 volunteers (15 females; mean age 29.5 [SD 5.1] years; mean height 1.7 [SD 0.8] m; mean weight 67.5 [SD 8.7] kg; mean maximal oxygen uptake 53.4 [SD 6.8] mL/min·kg) performed 7 activities—ranging in intensity from sitting to playing floorball—in a semistructured indoor environment for 10 min each, with 2-min breaks in between. These activities were performed while wearing the Polar Vantage M wristwatch and the MetaMax 3B spirometer.

Results:

After EE estimation, a mean (SD) of 69.1 (42.7) kcal and 71.4 (37.8) kcal per 10-min activity were reported for the MetaMax 3B and the Polar Vantage, respectively, with a strong correlation of r=0.892 (P<.001). The systematic bias was 2.3 kcal (3.3%), with 37.8 kcal limits of agreement. The lowest mean absolute percentage errors were reported during the sitting and reading activities (9.1%), and the highest error rates during household chores (31.4%). On average, 59.5% of the mean EE values obtained by the Polar Vantage were within ±20% of accuracy when compared with the MetaMax 3B. The activity intensity quantified by perceived exertion (odds ratio [OR] 2.028; P<.001) and wrist circumference (OR −1.533; P=.03) predicted 29% of the error rates within the Polar Vantage.

Conclusions:

The Polar Vantage has a statistically moderate-to-good accuracy in EE estimation that is activity dependent. During sitting and reading activities, the EE estimation is very good, whereas during nonsteady activities that require wrist and arm movement, the EE accuracy is only moderate. However, compared with other available wrist-worn EE monitors, the Polar Vantage can be recommended, as it performs among the best.


 Citation

Please cite as:

Gilgen-Ammann R, Schweizer T, Wyss T

Accuracy of the Multisensory Wristwatch Polar Vantage's Estimation of Energy Expenditure in Various Activities: Instrument Validation Study

JMIR Mhealth Uhealth 2019;7(10):e14534

DOI: 10.2196/14534

PMID: 31579020

PMCID: 6777286

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.