Accepted for/Published in: JMIR Rehabilitation and Assistive Technologies

Date Submitted: Apr 22, 2019
Date Accepted: Sep 26, 2019

The final, peer-reviewed published version of this preprint can be found here:

Wrist-Based Accelerometers and Visual Analog Scales as Outcome Measures for Shoulder Activity During Daily Living in Patients With Rotator Cuff Tendinopathy: Instrument Validation Study

Larrivée S, Balg F, Léonard G, Bédard S, Tousignant M, Boissy P

Wrist-Based Accelerometers and Visual Analog Scales as Outcome Measures for Shoulder Activity During Daily Living in Patients With Rotator Cuff Tendinopathy: Instrument Validation Study

JMIR Rehabil Assist Technol 2019;6(2):e14468

DOI: 10.2196/14468

PMID: 31793896

PMCID: 6918212

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Wrist-Based Accelerometers and Visual Analog Scales as Outcome Measures for Shoulder Activity During Daily Living in Patients With Rotator Cuff Tendinopathy: Instrument Validation Study

  • Samuel Larrivée; 
  • Frédéric Balg; 
  • Guillaume Léonard; 
  • Sonia Bédard; 
  • Michel Tousignant; 
  • Patrick Boissy

Background:

Shoulder pain secondary to rotator cuff tendinopathy affects a large proportion of patients in orthopedic surgery practices. Corticosteroid injections are a common intervention proposed for these patients. The clinical evaluation of a response to corticosteroid injections is usually based only on the patient’s self-evaluation of his function, activity, and pain by multiple questionnaires with varying metrological qualities. Objective measures of upper extremity functions are lacking, but wearable sensors are emerging as potential tools to assess upper extremity function and activity.

Objective:

This study aimed (1) to evaluate and compare test-retest reliability and sensitivity to change of known clinical assessments of shoulder function to wrist-based accelerometer measures and visual analog scales (VAS) of shoulder activity during daily living in patients with rotator cuff tendinopathy convergent validity and (2) to determine the acceptability and compliance of using wrist-based wearable sensors.

Methods:

A total of 38 patients affected by rotator cuff tendinopathy wore wrist accelerometers on the affected side for a total of 5 weeks. Western Ontario Rotator Cuff (WORC) index; Short version of the Disability of the Arm, Shoulder, and Hand questionnaire (QuickDASH); and clinical examination (range of motion and strength) were performed the week before the corticosteroid injections, the day of the corticosteroid injections, and 2 and 4 weeks after the corticosteroid injections. Daily Single Assessment Numeric Evaluation (SANE) and VAS were filled by participants to record shoulder pain and activity. Accelerometer data were processed to extract daily upper extremity activity in the form of active time; activity counts; and ratio of low-intensity activities, medium-intensity activities, and high-intensity activities.

Results:

Daily pain measured using VAS and SANE correlated well with the WORC and QuickDASH questionnaires (r=0.564-0.815) but not with accelerometry measures, amplitude, and strength. Daily activity measured with VAS had good correlation with active time (r=0.484, P=.02). All questionnaires had excellent test-retest reliability at 1 week before corticosteroid injections (intraclass correlation coefficient [ICC]=0.883-0.950). Acceptable reliability was observed with accelerometry (ICC=0.621-0.724), apart from low-intensity activities (ICC=0.104). Sensitivity to change was excellent at 2 and 4 weeks for all questionnaires (standardized response mean=1.039-2.094) except for activity VAS (standardized response mean=0.50). Accelerometry measures had low sensitivity to change at 2 weeks, but excellent sensitivity at 4 weeks (standardized response mean=0.803-1.032).

Conclusions:

Daily pain VAS and SANE had good correlation with the validated questionnaires, excellent reliability at 1 week, and excellent sensitivity to change at 2 and 4 weeks. Daily activity VAS and accelerometry-derived active time correlated well together. Activity VAS had excellent reliability, but moderate sensitivity to change. Accelerometry measures had moderate reliability and acceptable sensitivity to change at 4 weeks.


 Citation

Please cite as:

Larrivée S, Balg F, Léonard G, Bédard S, Tousignant M, Boissy P

Wrist-Based Accelerometers and Visual Analog Scales as Outcome Measures for Shoulder Activity During Daily Living in Patients With Rotator Cuff Tendinopathy: Instrument Validation Study

JMIR Rehabil Assist Technol 2019;6(2):e14468

DOI: 10.2196/14468

PMID: 31793896

PMCID: 6918212

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.