Maintenance Notice

Due to necessary scheduled maintenance, the JMIR Publications website will be unavailable from Wednesday, July 01, 2020 at 8:00 PM to 10:00 PM EST. We apologize in advance for any inconvenience this may cause you.

Who will be affected?

Accepted for/Published in: Journal of Medical Internet Research

Date Submitted: Dec 15, 2018
Date Accepted: Feb 9, 2020

The final, peer-reviewed published version of this preprint can be found here:

Real-Time Monitoring of Blood Pressure Using Digitalized Pulse Arrival Time Calculation Technology for Prompt Detection of Sudden Hypertensive Episodes During Laryngeal Microsurgery: Retrospective Observational Study

Park YS, Kim SH, Lee YS, Choi SH, Ku SW, Hwang GS

Real-Time Monitoring of Blood Pressure Using Digitalized Pulse Arrival Time Calculation Technology for Prompt Detection of Sudden Hypertensive Episodes During Laryngeal Microsurgery: Retrospective Observational Study

J Med Internet Res 2020;22(5):e13156

DOI: 10.2196/13156

PMID: 32412413

PMCID: 7260662

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Real-Time Monitoring of Blood Pressure Using Digitalized Pulse Arrival Time Calculation Technology for Prompt Detection of Sudden Hypertensive Episodes During Laryngeal Microsurgery: Retrospective Observational Study

  • Yong-Seok Park; 
  • Sung-Hoon Kim; 
  • Yoon Se Lee; 
  • Seung-Ho Choi; 
  • Seung-Woo Ku; 
  • Gyu-Sam Hwang

Background:

Laryngeal microsurgery (LMS) is often accompanied by a sudden increase in blood pressure (BP) during surgery because of stimulation around the larynx. This sudden change in the hemodynamic status is not immediately reflected in a casual cuff-type measurement that takes intermittent readings every 3 to 5 min.

Objective:

This study aimed to investigate the potential of pulse arrival time (PAT) as a marker for a BP surge, which usually occurs in patients undergoing LMS.

Methods:

Intermittent measurements of BP and electrocardiogram (ECG) and photoplethysmogram (PPG) signals were recorded during LMS. PAT was defined as the interval between the R-peak on the ECG and the maximum slope on the PPG. Mean PAT values before and after BP increase were compared. PPG-related parameters and the correlations between changes in these variables were calculated.

Results:

BP surged because of laryngoscopic manipulation (mean systolic BP [SBP] from 115.3, SD 21.4 mmHg, to 159.9, SD 25.2 mmHg; P<.001), whereas PAT decreased significantly (from mean 460.6, SD 51.9 ms, to 405.8, SD 50.1 ms; P<.001) in most of the cases. The change in SBP showed a significant correlation with the inverse of the PAT (r=0.582; P<.001). Receiver-operating characteristic curve analysis indicated that an increase of 11.5% in the inverse of the PAT could detect a 40% increase in SBP, and the area under the curve was 0.814.

Conclusions:

During LMS, where invasive arterial catheterization is not always possible, PAT shows good correlation with SBP and may, therefore, have the potential to identify abrupt BP surges during laryngoscopic manipulations in a noninvasive manner.

Clinicaltrial:


 Citation

Please cite as:

Park YS, Kim SH, Lee YS, Choi SH, Ku SW, Hwang GS

Real-Time Monitoring of Blood Pressure Using Digitalized Pulse Arrival Time Calculation Technology for Prompt Detection of Sudden Hypertensive Episodes During Laryngeal Microsurgery: Retrospective Observational Study

J Med Internet Res 2020;22(5):e13156

DOI: 10.2196/13156

PMID: 32412413

PMCID: 7260662

Per the author's request the PDF is not available.