Design of Improved Intertrochanteric Fracture Treatment (DRIFT) Study: Protocol for Biomechanical Testing and Finite Element Analysis of Stable and Unstable Intertrochanteric Fractures Treated With Intramedullary Nailing or Dynamic Compression Screw

Panagopoulos A, Kyriakopoulos G, Anastopoulos G, Megas P, Kourkoulis SK

Design of Improved Intertrochanteric Fracture Treatment (DRIFT) Study: Protocol for Biomechanical Testing and Finite Element Analysis of Stable and Unstable Intertrochanteric Fractures Treated With Intramedullary Nailing or Dynamic Compression Screw

JMIR Res Protoc 2019;8(7):e12845

DOI: 10.2196/12845

PMID: 31322133

PMCID: 6670281

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Design of Improved Intertrochanteric Fracture Treatment (DRIFT) Study: Protocol for Biomechanical Testing and Finite Element Analysis of Stable and Unstable Intertrochanteric Fractures Treated With Intramedullary Nailing or Dynamic Compression Screw

  • Andreas Panagopoulos; 
  • Georgios Kyriakopoulos; 
  • Georgios Anastopoulos; 
  • Panagiotis Megas; 
  • Stavros K Kourkoulis

Background:

Intertrochanteric hip fractures rank in the top 10 of all impairments worldwide in terms of loss in disability-adjusted years for people aged older than 60 years. The type of surgery is usually carried out with dynamic hip screw (DHS) devices or cephalomedullary nails (CMN). Cut-out of the hip screw is considered the most frequent mechanical failure for all implants with an estimated incidence ranging from 2% to 16.5%; this entails both enhancing our understanding of the prognostic factors of cut-out and improving all aspects of intertrochanteric fracture treatment.

Objective:

The Design of Improved Intertrochanteric Fracture Treatment (DRIFT) study’s main objective is to provide intertrochanteric fracture treatment expertise, requirements and specifications, clinical relevance, and validation to improve treatment outcomes by developing a universal algorithm for designing patient- and fracture-oriented treatment. The hypothesis to be tested is that a more valgus reduction angle and implants of higher angles will lead to a more favorable biomechanical environment for fracture healing—that is, higher compressive loads at the fracture site with lower shear loads at the hip screw femoral head interface. A new implant with enhanced biomechanical and technical characteristics will be designed and fabricated; in addition, an integrated design and optimization platform based on computer-aided design tools and topology optimization modules will be developed.

Methods:

To test this hypothesis, a biomechanical study comprising experimental loading of synthetic femora (Sawbones Inc) and finite element analysis (FEA) will be conducted. Detailed FEA of existing implants (DHS and CMN) implemented in different clinical cases under walking conditions will be performed to derive the stress and strain fields developed at the implant-bone system and identify critical scenarios that could lead to failure of therapy. These models would be validated against instrumented mechanical tests using strain gages and a digital image correlation process.

Results:

After testing, geometric drawbacks of existing implants will be fully recognized, and geometric characteristics will be correlated with critical failure scenarios. The last step would be the numeric design, computer-aided design (using FEA codes and design packages), and optimization of the new proposed implant with regard to improved biomechanical surgical technique and enhanced mechanical performance that will reduce the possibility for critical failure scenarios.

Conclusions:

The optimization of the biomechanical behavior of the fracture-osteosynthesis model by the application of the ideal reduction angle and implant is expected to have a positive effect to the rate of mechanical failure and, subsequently, the healing rates, morbidity, and mortality in this fragile patient group.

International Registered Report:

DERR1-10.2196/12845


 Citation

Please cite as:

Panagopoulos A, Kyriakopoulos G, Anastopoulos G, Megas P, Kourkoulis SK

Design of Improved Intertrochanteric Fracture Treatment (DRIFT) Study: Protocol for Biomechanical Testing and Finite Element Analysis of Stable and Unstable Intertrochanteric Fractures Treated With Intramedullary Nailing or Dynamic Compression Screw

JMIR Res Protoc 2019;8(7):e12845

DOI: 10.2196/12845

PMID: 31322133

PMCID: 6670281

Per the author's request the PDF is not available.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.