Evaluation of an Anesthesia Dashboard Functional Model Based on a Manufacturer-Independent Communication Standard: Comparative Feasibility Study

Ohligs M, Pereira C, Voigt V, Koeny M, Janß A, Rossaint R, Czaplik M

Evaluation of an Anesthesia Dashboard Functional Model Based on a Manufacturer-Independent Communication Standard: Comparative Feasibility Study

JMIR Hum Factors 2019;6(2):e12553

DOI: 10.2196/12553

PMID: 31042150

PMCID: 6658249

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Evaluation of an Anesthesia Dashboard Functional Model Based on a Manufacturer-Independent Communication Standard: Comparative Feasibility Study

  • Marian Ohligs; 
  • Carina Pereira; 
  • Verena Voigt; 
  • Marcus Koeny; 
  • Armin Janß; 
  • Rolf Rossaint; 
  • Michael Czaplik

Background:

Current anesthesia workspaces consist of several technical devices, such as patient monitors, anesthesia machines, among others. Commonly, they are produced by different manufacturers; thus, they differ in terms of their modus operandi, user interface, and representation of alarms. Merging the information from these devices using a single joint protocol and displaying it in a single graphical user interface could lead to a general improvement in perioperative management. For this purpose, the recently approved and published Institute of Electrical and Electronics Engineers 11073 service-oriented device connectivity standard was implemented.

Objective:

This paper aims to develop and then evaluate an anesthesia workstation (ANWS) functional model in terms of usability, fulfillment of clinical requirements, and expected improvements in patient safety.

Methods:

To compare the self-developed ANWS with the conventional system, a pilot observational study was conducted at the University Hospital Aachen, Germany. A total of 5 anesthesiologists were asked to perform different tasks using the ANWS and then the conventional setup. For evaluation purposes, response times were measured and an interaction-centered usability test with an eye-tracking system was carried out. Finally, the subjects were asked to fill in a questionnaire in order to measure user satisfaction.

Results:

Response times were significantly higher when using the ANWS, but decreased considerably after one repetition. Furthermore, usability was rated as excellent (≥95) according to the System Usability Scale score, and the majority of clinical requirements were met.

Conclusions:

In general, the results were highly encouraging, considering that the ANWS was only a functional model, as well as the lack of training of the participants. However, further studies are necessary to improve the universal user interface and the interplay of the various networked devices.


 Citation

Please cite as:

Ohligs M, Pereira C, Voigt V, Koeny M, Janß A, Rossaint R, Czaplik M

Evaluation of an Anesthesia Dashboard Functional Model Based on a Manufacturer-Independent Communication Standard: Comparative Feasibility Study

JMIR Hum Factors 2019;6(2):e12553

DOI: 10.2196/12553

PMID: 31042150

PMCID: 6658249