Diagnostic Performance of a Smart Device With Photoplethysmography Technology for Atrial Fibrillation Detection: Pilot Study (Pre-mAFA II Registry)

Fan YY, Li YG, Li J, Cheng WK, Shan ZL, Wang YT, Guo YT

Diagnostic Performance of a Smart Device With Photoplethysmography Technology for Atrial Fibrillation Detection: Pilot Study (Pre-mAFA II Registry)

JMIR Mhealth Uhealth 2019;7(3):e11437

DOI: 10.2196/11437

PMID: 30835243

PMCID: 6423467

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Diagnostic Performance of a Smart Device With Photoplethysmography Technology for Atrial Fibrillation Detection: Pilot Study (Pre-mAFA II Registry)

  • Yong-Yan Fan; 
  • Yan-Guang Li; 
  • Jian Li; 
  • Wen-Kun Cheng; 
  • Zhao-Liang Shan; 
  • Yu-Tang Wang; 
  • Yu-Tao Guo

Background:

Atrial fibrillation (AF) is the most common sustained cardiac arrhythmia. The asymptomatic nature and paroxysmal frequency of AF lead to suboptimal early detection. A novel technology, photoplethysmography (PPG), has been developed for AF screening. However, there has been limited validation of mobile phone and smart band apps with PPG compared to 12-lead electrocardiograms (ECG).

Objective:

We investigated the feasibility and accuracy of a mobile phone and smart band for AF detection using pulse data measured by PPG.

Methods:

A total of 112 consecutive inpatients were recruited from the Chinese PLA General Hospital from March 15 to April 1, 2018. Participants were simultaneously tested with mobile phones (HUAWEI Mate 9, HUAWEI Honor 7X), smart bands (HUAWEI Band 2), and 12-lead ECG for 3 minutes.

Results:

In all, 108 patients (56 with normal sinus rhythm, 52 with persistent AF) were enrolled in the final analysis after excluding four patients with unclear cardiac rhythms. The corresponding sensitivity and specificity of the smart band PPG were 95.36% (95% CI 92.00%-97.40%) and 99.70% (95% CI 98.08%-99.98%), respectively. The positive predictive value of the smart band PPG was 99.63% (95% CI 97.61%-99.98%), the negative predictive value was 96.24% (95% CI 93.50%-97.90%), and the accuracy was 97.72% (95% CI 96.11%-98.70%). Moreover, the diagnostic sensitivity, specificity, positive predictive value, negative predictive value, and accuracy of mobile phones with PPG for AF detection were over 94%. There was no significant difference after further statistical analysis of the results from the different smart devices compared with the gold-standard ECG (P>.99).

Conclusions:

The algorithm based on mobile phones and smart bands with PPG demonstrated good performance in detecting AF and may represent a convenient tool for AF detection in at-risk individuals, allowing widespread screening of AF in the population.

ClinicalTrial:

Chinese Clinical Trial Registry ChiCTR-OOC-17014138; http://www.chictr.org.cn/showproj.aspx?proj=24191 (Archived by WebCite at http://www.webcitation/76WXknvE6)


 Citation

Please cite as:

Fan YY, Li YG, Li J, Cheng WK, Shan ZL, Wang YT, Guo YT

Diagnostic Performance of a Smart Device With Photoplethysmography Technology for Atrial Fibrillation Detection: Pilot Study (Pre-mAFA II Registry)

JMIR Mhealth Uhealth 2019;7(3):e11437

DOI: 10.2196/11437

PMID: 30835243

PMCID: 6423467

Per the author's request the PDF is not available.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.