Accepted for/Published in: JMIR Cardio
Date Submitted: Jun 18, 2018
Open Peer Review Period: Jun 21, 2018 - Aug 16, 2018
Date Accepted: Oct 26, 2018
(closed for review but you can still tweet)
Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.
Feasibility of Telemonitoring Blood Pressure in Patients With Kidney Disease (Oxford Heart and Renal Protection Study-1): Observational Study
Background:
Blood pressure (BP) is a key modifiable risk factor for patients with chronic kidney disease (CKD), with current guidelines recommending strict control to reduce the risk of progression of both CKD and cardiovascular disease. Trials involving BP lowering require multiple visits to achieve target BP, which increases the costs of such trials, and in routine care, BP measured in the clinic may not accurately reflect the usual BP.
Objective:
We sought to assess whether a telemonitoring system for BP (using a Bluetooth-enabled BP machine that could transmit BP measurements to a tablet device installed with a bespoke app to guide the measurement of BP and collect questionnaire data) was acceptable to patients with CKD and whether patients would provide sufficient BP readings to assess variability and guide treatment.
Methods:
A total of 25 participants with CKD were trained to use the telemonitoring equipment and asked to record BP daily for 30 days, attend a study visit, and then record BP on alternate days for the next 60 days. They were also offered a wrist-worn applanation tonometry device (BPro) which measures BP every 15 minutes over a 24-hour period. Participants were given questionnaires at the 1- and 3-month time points; the questionnaires were derived from the System Usability Scale and Technology Acceptance Model. All eligible participants completed the study.
Results:
Mean participant age was 58 (SD 11) years, and mean estimated glomerular filtration rate was 36 (SD 13) mL/min/1.73m2. 13/25 (52%) participants provided >90% of the expected data and 18/25 (72%) provided >80% of the expected data. The usability of the telemonitoring system was rated highly, with mean scores of 84.9/100 (SE 2.8) after 30 days and 84.2/100 (SE 4.1) after 90 days. The coefficient of variation for the variability of systolic BP telemonitoring was 9.4% (95% CI 7.8-10.9) compared with 7.9% (95% CI 6.4-9.5) for the BPro device,
Conclusions:
Telemonitoring is acceptable for patients with CKD and provides sufficient data to inform titration of antihypertensive therapies in either a randomized trial setting (comparing BP among different targets) or routine clinical practice. Such methods could be employed in both scenarios and reduce costs currently associated with such activities.
ClinicalTrial:
International Standard Randomized Controlled Trial Number ISRCTN13725286; http://www.isrctn.com/ISRCTN13725286 (Archived by WebCite at http://www.webcitation.org/74PAX51Ji).
Citation
Per the author's request the PDF is not available.
Copyright
© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.