The Eye-Pod: A multi-purpose modular system for cataract monitoring

  • Clare Cheng; 
  • Rubin How Sheng Yong; 
  • Pamelia Mei En Leong; 
  • Vivien Sin Yi Lim; 
  • Victor Teck Chang Koh; 
  • David Ziyou Chen

ABSTRACT

Background:

Cataract is the leading cause of blindness in the Asia-Pacific region, where demographic ageing, limited ophthalmologist density, and constrained outpatient capacity create a growing need for scalable community surveillance. Patients below surgical thresholds require prolonged monitoring, yet no self-operated multimodal cataract surveillance model has been evaluated in this region. We developed the Eye-Pod - a modular self-operated booth integrating visual acuity testing, rebound tonometry, anterior segment imaging, and fundus photography. We conducted a proof-of-concept validation study to evaluate its concordance with clinical surgical recommendations, individual device performance, and patient acceptability.

Objective:

To evaluate the concordance of a modular, self-operated surveillance booth (Eye-Pod) with clinical surgical recommendations, and to assess individual device performance and patient acceptability, in a prospective, single-centre study of 150 participants (300 eyes), aged 40–80, with no previous cataract or intraocular surgery, recruited from the National University Hospital, Singapore.

Methods:

Participants completed self-administered visual acuity testing, tonometry (iCare), anterior segment imaging (MIDAS), and fundus photography within the Eye-Pod. Images were asynchronously graded by a consultant ophthalmologist across two independent rounds. Concordance with clinical recommendations for cataract surgery was assessed using Cohen's kappa and PABAK. Generalised estimating equations identified factors associated with concordance. Usability was evaluated with the System Usability Scale (SUS) and After-Scenario Questionnaire (ASQ).

Results:

293 eyes (97.7%) were gradable. Overall concordance was slight (κ=0.16, PABAK=0.31), with accuracy 66.2%, sensitivity 44.2%, specificity 73.0%, and NPV 81.0%. Visual acuity concordance (aOR 2.37, p=0.002) and clinically significant cataract (aOR 0.37, p=0.002) were independent predictors of concordance on multivariable GEE. Mean SUS score was 64.2 (SD 17.2); 61 participants (40.7%) achieved good usability. Male gender was the only independent predictor of good usability (OR 2.57, 95% CI 1.31–5.03, p=0.006). Age (≥70 years) and hypertension were independently associated with lower VA concordance.

Conclusions:

The Eye-Pod demonstrates proof of concept as a self-administered cataract surveillance platform, with high specificity and acceptable usability. However, overall concordance was slight, and performance was attenuated in older patients and those with clinically significant cataract, indicating that further refinement is needed before community deployment


 Citation

Please cite as:

Cheng C, Yong RHS, Leong PME, Lim VSY, Koh VTC, Chen DZ

The Eye-Pod: A multi-purpose modular system for cataract monitoring

JMIR Preprints. 29/08/2026:110558

DOI: 10.2196/preprints.110558

URL: https://preprints.jmir.org/preprint/110558

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.