The final, peer-reviewed published version of this preprint can be found here:

A Hazard Analysis of Class I Recalls of Infusion Pumps

Gao X, Wen Q, Duan X, Jin W, Tang X, Zhong L, Xia S, Feng H, Zhong D

A Hazard Analysis of Class I Recalls of Infusion Pumps

JMIR Hum Factors 2019;6(2):e10366

DOI: 10.2196/10366

PMID: 31066695

PMCID: 6524450

A Hazard Analysis of Class I Recalls of Infusion Pump

  • Xuemei Gao; 
  • Qiang Wen; 
  • Xiaolian Duan; 
  • Wei Jin; 
  • Xiaohong Tang; 
  • Ling Zhong; 
  • Shitao Xia; 
  • Hailing Feng; 
  • Daidi Zhong

ABSTRACT

Background:

The adverse event report of medical devices is one of the post-market surveillance tools for regulators to monitor device performance, detect potential device-related safety issues, and contribute to benefit-risk assessments of these products. Along with the development of the related technologies and market, the amount of adverse events keeps increasing, which results in the need for efficient tools that help to analyze the adverse events monitoring data and to identify the risk signals.

Objective:

To establish a hazard classification framework of the medical devices, and to apply it over practical adverse event data regarding infusion pumps. Subsequently, to analyze the risks of infusion pumps, and to provide reference for the risk management of this type of device.

Methods:

The authors defines a general hierarchical classification of medical device hazards. This classification is combined with the Trace Intersecting Theory to form a human-machine-environment interaction model. Such model is applied to the dataset of 2001~2017 class Ⅰ infusion pump recalls extracted from FDA website. This dataset does not include the cases caused by illegal factors, in order to reflect the risk signals of this type of device.

Results:

The proposed model is leveraged in the hazard analysis over 70 cases of class I infusion pump recalls by FDA. According to the analytical results, the “infusion pump dose not infuse accurate dosage (over or under delivery of fluid)” is identified to be an important source of product technical risk. The Energy hazard is the major hazard form for infusion pumps. The product component failure is the main direct cause for the studied cases.

Conclusions:

The proposed human-machine-environment interaction model, when applied to adverse event data, can help to identify the hazard forms and direct causes of medical device adverse events.


 Citation

Please cite as:

Gao X, Wen Q, Duan X, Jin W, Tang X, Zhong L, Xia S, Feng H, Zhong D

A Hazard Analysis of Class I Recalls of Infusion Pumps

JMIR Hum Factors 2019;6(2):e10366

DOI: 10.2196/10366

PMID: 31066695

PMCID: 6524450

Per the author's request the PDF is not available.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.