Validation of heart rate- and electrodermal-related signals from commercial grade wearable smartwatch and rings in experimentally induced stress: a laboratory-based study
ABSTRACT
Background:
Commercial wearable devices such as smartwatches and smart rings have potential for remote monitoring of stress physiology in health research. These devices are commonly equipped with photoplethysmogram (PPG) sensors to measure heart rate (HR) and heart rate variability (HRV) and/or electrodermal activity (EDA) sensors for skin conductance, indicators of stress. Data obtained from the sensors should be validated against measurements from gold-standard laboratory equipment (Mindware) before being used in clinical research.
Objective:
The study aims to evaluate the sensitivity of measurements obtained from four wearable devices (Nowatch X, Empatica EmbracePlus, Nuanic, and Oura Ring) in controlled lab-based stress-induction tasks, as well as the correspondence between the measurements from the wearable and lab devices.
Methods:
62 healthy participants completed four stress induction tasks while wearing the above mentioned wearable devices, along with gold-standard EDA and electrocardiogram (ECG) electrodes from Mindware Technology. HR, HRV, EDA and EDA-derived metrics were calculated and compared against the lab recordings during the stress inducing tasks and interleaved rest periods. We used repeated-measure ANOVAs to assess within-devices change during rest and stress protocols. Linear mixed-effect models, Bland-Altman plots, and Pearson correlation coefficients were used to evaluate the strength of association and agreement between measurements obtained from the wearable devices and the laboratory device.
Results:
Laboratory stress-induction tasks elicited subjective stress and physiological responses as measured by both wearable and lab devices. Nuanic’s EDA showed strong association with laboratory EDA measures at a signal level (r = 0.78). Nowatch HR and MeanNN closely matched Mindware ECG measures (r = 0.87 and r = 0.85), whereas Standard Deviation of Normal-to-Normal intervals (SDNN) and Root Mean Square of Successive Differences (RMSSD) showed weaker associations (r = 0.55 and r = 0.42). Empatica also demonstrated a strong association with Mindware (r = 0.85), while Oura Ring’s processed HR did not correspond well (r = 0.68) with Mindware. Both Empatica and Oura Ring had weaker correspondence in HRV compared with Mindware ECG-derived HRV (RMSSD; r = 0.36 and r = 0.41). Bland-Altman analyses showed relatively small mean biases for most measures. Wearable-derived HR tended to be slightly lower than ECG-derived HR, whereas the direction of bias varied across EDA and HRV measures. Wearable SCR amplitude and wearable HR measures showed narrower limits of agreement and lower mean absolute errors, whereas HRV metrics generally showed wider limits of agreement and larger mean absolute errors.
Conclusions:
Overall, the wearable devices showed good sensitivity to experimentally induced stress and strong correspondence with laboratory-based HR and EDA measures in a controlled environment, whereas HRV indices were less consistent. Further validation under ambulatory and free-living conditions is needed to determine the performance of these devices in real-world monitoring applications.
Citation
Request queued. Please wait while the file is being generated. It may take some time.
Copyright
© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.