Reliability and Safety of Smartwatch Blood Pressure and Oxygen Saturation Measurements in Older Adults: Instrument Validation Study

Madeira R, Esteves D, Neiva H, Maia A, Rondão C, Pinto N, Vercelli A, Pato MV

Reliability and Safety of Smartwatch Blood Pressure and Oxygen Saturation Measurements in Older Adults: Instrument Validation Study

JMIR Form Res 2026;10:e81955

DOI: 10.2196/81955

Reliability and Safety of Smartwatch Blood Pressure and Oxygen Saturation Measurements in Older Adults: Original paper

  • Ricardo Madeira; 
  • Dulce Esteves; 
  • Henrique Neiva; 
  • Adriana Maia; 
  • Catarina Rondão; 
  • Nuno Pinto; 
  • Alessandro Vercelli; 
  • Maria Vaz Pato

ABSTRACT

Background:

Hypertension is a significant risk factor for cardiovascular diseases and premature mortality, with its prevalence increasing due to population aging and lifestyle factors. Accurate measurement of blood pressure and arterial oxygen saturation is crucial for disease prevention and monitoring, and wearable devices have emerged as a promising alternative. However, their clinical reliability requires validation, particularly in older populations

Objective:

The aim of this research is to evaluate and compare the measurement of blood pressure and arterial oxygen saturation in elderly people using the Smartwatch in comparison with reference devices.

Methods:

We recruited 50 subjects aged between 50 and 89 years (70.60±12.03 years), including 34 were women and 16 were men. Three blood pressure measurements were taken simultaneously using the Smartwatch and an Ambulatory Blood pressure monitoring (reference device). Arterial oxygen saturation was measured simultaneously using the Smartwatch and the oximeter. The T-test for independent samples was used to compare variables and the intraclass correlation coefficient to verify the correlation.

Results:

Considering the correlation between blood pressure measured by the Smartwatch and Ambulatory Blood Pressure Monitoring, the systolic blood pressure variable in the three measurements taken is rated Good (ICC≤0.87 and ICC≥0.90). In the diastolic blood pressure variable, the correlation varies between Good and Excellent (ICC≤0.82 and ICC≥0.91). The independent-sample T-test between the blood pressure samples on the systolic blood pressure variable in the three measurements showed no differences (P=0.64; P=0.74 and P=0.86). In the diastolic blood pressure variable, there was also no significance (P=0.80; P=0.51 and P=0.84). The correlation between the Arterial Oxygen Saturation measured by the Smartwatch and the Oximeter in the Arterial Oxygen Saturation variable was moderate (ICC=0.68). The independent-samples T-test between the Arterial Oxygen Saturation samples showed no differences (P=0.67).

Conclusions:

Our study showed that the Smartwatch is capable of reliably measuring resting blood pressure and arterial oxygen saturation in an elderly population.


 Citation

Please cite as:

Madeira R, Esteves D, Neiva H, Maia A, Rondão C, Pinto N, Vercelli A, Pato MV

Reliability and Safety of Smartwatch Blood Pressure and Oxygen Saturation Measurements in Older Adults: Instrument Validation Study

JMIR Form Res 2026;10:e81955

DOI: 10.2196/81955

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.