Accuracy, Reproducibility, and Responsiveness to Treatment of Home Spirometry in Cystic Fibrosis: Multicenter, Retrospective, Observational Study

Oppelaar MC, van Helvoort H, Bannier MA, Reijers MH, van der Vaart H, van der Meer R, Altenburg J, Conemans L, Rottier BL, Nuijsink M, van den Wijngaart LS, Merkus PJ, Roukema J

Accuracy, Reproducibility, and Responsiveness to Treatment of Home Spirometry in Cystic Fibrosis: Multicenter, Retrospective, Observational Study

J Med Internet Res 2024;26:e60892

DOI: 10.2196/60892

PMID: 39626236

PMCID: 11653036

Accuracy, reproducibility, and responsiveness to treatment of home spirometry in cystic fibrosis: A Multicentre Retrospective Observational Study.

  • Martinus C. Oppelaar; 
  • Hanneke van Helvoort; 
  • Michiel A.G.E. Bannier; 
  • Monique H.E. Reijers; 
  • Hester van der Vaart; 
  • Renske van der Meer; 
  • Josje Altenburg; 
  • Lennart Conemans; 
  • Bart L. Rottier; 
  • Marianne Nuijsink; 
  • Lara S. van den Wijngaart; 
  • Peter J.F.M. Merkus; 
  • Jolt Roukema

ABSTRACT

Background:

Portable spirometers are increasingly used to measure lung function at home, but doubts about the accuracy of these devices persist. These doubts stand in the way of the digital transition of chronic respiratory disease care, hence there is a need to address the accuracy of home spirometry in routine care across multiple settings and ages.

Objective:

We aimed to assess accuracy, reproducibility and responsiveness to treatment of home spirometry in long-term paediatric and adult CF care.

Methods:

Retrospective observational study in five Dutch CF centres. Home spirometry outcomes (Forced Expiratory Volume in one second, FEV1; Forced Vital Capacity, FVC) for 601 anonymized users were collected during three years. For 81 users, data on clinic spirometry and elexacaftor/tezacaftor/ivacaftor (ETI) use were available. Accuracy was assessed using Bland-Altman plots. Intra-test reproducibility was assessed with ATS/ERS spirometry quality grades. Responsiveness was measured by %change in home spirometry outcomes after start of ETI.

Results:

No heteroscedasticity was present in the 263 clinic-home spirometry pairs (n=81). Home spirometry was significantly lower than clinic spirometry (Mean ΔFEV1clinic-home 0.13 L; Mean ΔFVCclinic-home 0.20 L) across ages and over time. 5104 unique home measurements were graded. Grade E was given to 2435 tests as only one home measurement was performed. Of the remaining 2669 tests, 43.8% and 43.6% received grade A and B respectively. Median %change in FEV1 from baseline after initiation of ETI was 19.2% after 7-14 days and remained stable thereafter (n=33).

Conclusions:

Home spirometry is feasible but not equal to clinic spirometry. Home spirometry can confirm whether lung functions remain stable, but the context of a measurement and personal trends are more relevant than absolute outcomes.


 Citation

Please cite as:

Oppelaar MC, van Helvoort H, Bannier MA, Reijers MH, van der Vaart H, van der Meer R, Altenburg J, Conemans L, Rottier BL, Nuijsink M, van den Wijngaart LS, Merkus PJ, Roukema J

Accuracy, Reproducibility, and Responsiveness to Treatment of Home Spirometry in Cystic Fibrosis: Multicenter, Retrospective, Observational Study

J Med Internet Res 2024;26:e60892

DOI: 10.2196/60892

PMID: 39626236

PMCID: 11653036

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.