Insertable loop recorder implantation and follow-up by specialized electrophysiology nurses: safety, signal quality, and patient satisfaction
ABSTRACT
Background:
Insertable loop recorders (ILRs) are subcutaneous cardiac monitoring devices used in the diagnosis of arrhythmias and in the etiological investigation of syncope and cryptogenic stroke. Device miniaturization and procedural simplification have facilitated implantation in ambulatory settings. The involvement of specialized electrophysiology nurses may help meet rising clinical demand, although evidence in our setting remains limited.
Objective:
This study aimed to analyze implantation safety, detection signal quality, adverse events during follow-up, and patient satisfaction when the procedure is performed by specialized nurses.
Methods:
This prospective observational study (the CARDIONURSE study) was conducted in the Arrhythmia Unit of Hospital General Universitario Doctor Balmis between February 2022 and May 2023. Thirty-five patients with a clinical indication for ILR implantation were included. The procedure was performed by 2 specialized nurses under medical supervision. Sociodemographic, clinical, and technical implantation variables were collected, including detection signal amplitude (mV), together with adverse events and follow-up outcomes at 1 and 6 months. Patient satisfaction was assessed using a validated questionnaire at 4 time points. Descriptive statistical analysis was performed.
Results:
Mean age was 71.1 (SD 11.3) years, and 62.9% were women. The most frequent indication was syncope (77.1%), followed by suspected atrial fibrillation (AF) in cryptogenic stroke (17.1%). The procedure was performed without complications in 85.7% of cases and lasted less than 5 minutes in 77.1%. Mean signal amplitude was 0.45 (SD 0.29) mV, with a visible P wave in 94.3% of patients. At 6 months, 85.7% had no wound-related adverse events. A diagnosis was obtained in 20% of cases. Overall, 97.1% of patients reported maximum or very high satisfaction with the intervention, and 100% with the care received.
Conclusions:
ILR implantation by specialized electrophysiology nurses in an ambulatory setting is a safe, fast, and technically adequate procedure, with excellent signal quality and high patient satisfaction. These results support the incorporation of specialized nursing as ILR implanters within the Spanish health care system. Clinical Trial: Not applicable; this is a prospective observational study, not a clinical trial.
Citation
Request queued. Please wait while the file is being generated. It may take some time.
Copyright
© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.