Semipermanent Smartphone Adapter for Microscopes: Design Demonstration and Workflow Testing Using a Slit-Lamp Biomicroscope

  • Chieh-Hung Yen; 
  • Guo-Quan Wang; 
  • Tommy Y. Lin; 
  • Chun-Hsiu Liu

Background:

A smartphone microscope adapter is an alternative to a permanently mounted camera, which often requires a specially designed light path as well as a computer system for data storage and network connectivity. A common disadvantage of these adapters is that they need to be repeatedly mounted and dismounted; this is inconvenient during serial observations.

Objective:

To develop and test a semipermanent microscope adapter.

Methods:

Our prototype was designed using three-dimensional (3D) computer-aided design software and was 3D printed for testing. Two tests were designed to quantify the workflow improvement by measuring the time required to complete specific movements, which simulated the actions of capturing photographs by using an adapter and a slit-lamp biomicroscope. Our prototype was the experimental adapter, and two commercially available products (Snapzoom and Carson HookUpz IS-100) were used as control or reference adapters. A single-shot test measured the average time required to complete a photographing cycle, in which a single photo was recorded. A multiple-shot test measured the average time required to complete cycles in which different numbers of photos were captured; thus, the time required specifically for the preparation phase of photographing was determined. Both tests were performed by two researchers, a third-year resident and visiting staff member.

Results:

In the single-shot test, the average time required using our prototype to complete a photographing cycle was significantly shorter than the average time required using the two commercially available products (p < 0.001). In the multiple-shot test, the linear regression analysis also showed the preparation phase of our prototype was significantly shorter than the preparation phase of the two commercially available products (p < 0.001).

Conclusions:

The semipermanent design may improve the workflow of photography using smartphone adapters, thereby reducing the time and effort required for preparing devices, particularly when making multiple consecutive observations.