Maintenance Notice

Due to necessary scheduled maintenance, the JMIR Publications website will be unavailable from Wednesday, July 01, 2020 at 8:00 PM to 10:00 PM EST. We apologize in advance for any inconvenience this may cause you.

Who will be affected?

Previously submitted to: JMIR Human Factors (no longer under consideration since Sep 17, 2026)

Date Submitted: Feb 23, 2026

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Increased radiation exposure after a CT software upgrade: insights from patient safety classification and workflow analysis

  • Md Shafiqur Rahman Jabin

ABSTRACT

Background:

Software upgrades in medical imaging systems are routinely implemented to improve functionality and performance, but they may also introduce unintended changes in system behaviour that affect patient safety. Evidence on how such incidents manifest within imaging workflows remains limited.

Objective:

This study aimed to analyze a software upgrade–related incident in computed tomography (CT) that resulted in unintended increases in radiation dose, using established patient safety and health information technology (HIT) frameworks.

Methods:

A de-identified incident report from a national incident reporting repository was analyzed using a combined deductive and inductive qualitative approach. Deductive analysis applied the International Classification for Patient Safety (ICPS), the Health Information Technology Classification System (HIT-CS), and the 18-step medical imaging process workflow. Inductive thematic analysis was conducted to identify contextual factors within the incident narrative.

Results:

The incident was classified as a technical HIT issue related to software functionality, with incorrect machine output resulting in increased radiation dose despite unchanged examination protocols. Under ICPS, the event was categorized as a medical device/system incident with communication and technology design as contributing factors. Workflow analysis identified image acquisition as the primary step where the incident manifested, with detection occurring during retrospective review. Thematic analysis highlighted limited visibility of software-induced changes and reliance on retrospective monitoring, informing the development of preventive and corrective strategies.

Conclusions:

This study demonstrates that software upgrade–related HIT incidents in CT imaging can lead to unintended increases in radiation dose during routine clinical use.


 Citation

Please cite as:

Rahman Jabin MS

Increased radiation exposure after a CT software upgrade: insights from patient safety classification and workflow analysis

JMIR Preprints. 23/02/2026:94009

DOI: 10.2196/preprints.94009

URL: https://preprints.jmir.org/preprint/94009

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.