Assessment of a Wearable Device to Measure 6-Minute Walk Distance in Older Adults: A Pilot Study

  • Justin J. Cheng; 
  • Taylor Johnson; 
  • Arash Naeim

ABSTRACT

Background:

The 6-minute walk test (6MWT) is used to assess functional capacity and has various clinical applications. However, its implementations can be resource-intensive, requiring sufficient staffing and logistical support. Emerging digital technologies, such as wearable devices, offer promising solutions to ease these operational challenges. One example is the Clario Opal V2C, which recently added 6MWT to its existing technology.

Objective:

Our pilot study sought to assess how the Opal measurement of 6MW distance (6MWD) compared to manual assessment and whether the sensors affected participants’ walking ability, as measured by 6MWD and a Likert-scale survey.

Methods:

Participants included older adults at two independent living facilities. The 6MWT was conducted twice – once with and once without the Opal sensors – with a 15-minute rest in between. Manual measurement was conducted using a measuring wheel and compared to Opal recorded distance by paired t-test. Significance was defined as P <.05. Subjects were also asked about their thoughts on the Opal sensors through a Likert-scale survey.

Results:

The pilot study included 15 older participants (mean age 76.7 years). Opal recorded 6MWD was on average 12.1 ± 16.6 meters less than manual measurement (P=.01). There was no difference in 6MWD with or without the Opal sensors (mean 7.41 ± 27.7, P=.32). Participants had favorable views of the Opal sensors regarding its look and comfort, with all indicating “agree” or “strongly agree” that it did not affect their walking.

Conclusions:

There was a difference between manual and Opal measurement of 6MWD, but the sensors do not appear to impact walking distance or ability. Its use in larger clinical studies can potentially alleviate staff and time constraints in research that assesses 6MWD as an outcome measure.


 Citation

Please cite as:

Cheng JJ, Johnson T, Naeim A

Assessment of a Wearable Device to Measure 6-Minute Walk Distance in Older Adults: A Pilot Study

JMIR Preprints. 19/05/2025:77060

DOI: 10.2196/preprints.77060

URL: https://preprints.jmir.org/preprint/77060

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.