Fitbit’s accuracy to measure short bouts of physical activity and sedentary behavior: validation, sensitivity and specificity study

  • Julie Delobelle; 
  • Elien Lebuf; 
  • Delfien Van Dyck; 
  • Sofie Compernolle; 
  • Michael Janek; 
  • Femke De Backere; 
  • Tomas Vetrovsky

ABSTRACT

Background:

Adopting a physically active lifestyle results in substantial health benefits. However, many adults and older adults are insufficiently physically active and engage in high levels of (SB). Just-in-time adaptive interventions (JITAIs) are promising in targeting these health behaviors by providing real-time supportive prompts during opportunistic moments. Developing effective JITAIs requires better insight into the dynamic determinants of PA and SB, which can be obtained through event-based ecological momentary assessment (EMA). In both JITAIs and event-based EMA research, real-time monitoring of participants’ behavior is needed to enable the detection of the events of interest (e.g., 5 minutes of sustained walking or 30 minutes of prolonged sitting).

Objective:

To ensure Fitbits are suitable devices for use in event-based EMA studies and JITAIs, we explored the concurrent validity, accuracy and inter-device agreement of two Fitbit models (Inspire 2 and Ionic) in measuring short bouts of stepping and SB among free-living adults and older adults.

Methods:

Thirty-seven adults (18-65y) and 32 older adults (65+) from two countries (Belgium and the Czech Republic) wore four devices simultaneously for three consecutive days. In addition to two Fitbit models at the non-dominant wrist, they wore an ActiGraph GT3X+ at the hip and an ActivPAL at the thigh as reference devices for step count and SB, respectively. To investigate the concurrent validity for measuring step count during sustained walking (≥ 60 steps/min), the mean error and mean absolute error were calculated for 1-, 5-, 10-, and 20-minute epochs. In addition, concordance correlation coefficients were calculated and Bland-Altman plots were constructed. To explore Fitbit’s accuracy in detecting short bouts of stepping, the sensitivities and specificities were calculated for various thresholds (60 to 120 steps/minute) and bout durations (1 to 20 min). Finally, to define Fitbit’s optimal thresholds to capture prolonged bouts of sitting (1 to 60 min), receiver operating characteristics curves were constructed.

Results:

The Fitbit appeared to be valid for measuring steps on a short time scale (i.e., 1, 20, 30 and 60 minutes) compared to the ActiGraph. Except for stepping above 120 steps/min in older adults, both Fitbits detected stepping bouts of various durations in both adults and older adults with sensitivities and specificities greater than 87% and 97%, respectively. In addition, optimal cut-off values to detect prolonged bouts of sitting were identified, obtaining sensitivities and specificities greater than 93% and 89%, respectively.

Conclusions:

This paper provides practical considerations for using Fitbit devices in JITAIs and event-based EMA studies in adults and older adults. Considering Fitbits' reasonable accuracy in detecting short bouts of stepping and SB, they can be used to trigger JITAI prompts or EMA questionnaires following an event of interest (e.g., 5 minutes of sustained walking).


 Citation

Please cite as:

Delobelle J, Lebuf E, Van Dyck D, Compernolle S, Janek M, De Backere F, Vetrovsky T

Fitbit’s accuracy to measure short bouts of physical activity and sedentary behavior: validation, sensitivity and specificity study

JMIR Preprints. 24/11/2022:44568

DOI: 10.2196/preprints.44568

URL: https://preprints.jmir.org/preprint/44568

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.