Evaluating the influence of hematocrit: comparison of three POCT devices pursuant to ISO 15197 guidelines.

  • Matthes Kenning; 
  • Anselm Puchert; 
  • Eckhard Salzsieder

ABSTRACT

Background:

Maintaining normal blood glucose levels in diabetes therapy is fundamentally linked to precise and accurate blood glucose measurements. In the light of interfering quantities like various substances or hematocrit, these aspects are of utmost clinical importance within a standardized validation of blood glucose monitors for personal use as well as a continuous quality assessment.

Objective:

Continuous and independent quality assurance of aspects affecting a blood glucose monitoring system’s performance.

Methods:

The influence of the hematocrit on the performance of three blood glucose monitors was assessed in accordance with ISO 15197 and sponsor’s requirements.

Results:

Only one device showed acceptable deviations of ≤ 10 mg/dL / 10 % to the reference device across the entire range. The other devices showed minor to clinically relevant interferences, particularly in response to higher hematocrit values.

Conclusions:

Suitable BGM systems should be selected carefully by health care professionals and patients, especially with medical conditions known to interfere with blood glucose measurements.


 Citation

Please cite as:

Kenning M, Puchert A, Salzsieder E

Evaluating the influence of hematocrit: comparison of three POCT devices pursuant to ISO 15197 guidelines.

JMIR Preprints. 14/09/2021:33574

DOI: 10.2196/preprints.33574

URL: https://preprints.jmir.org/preprint/33574

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.