Validation of heart rate measurement of Fitbit Charge 4 and Xiaomi Mi Band 5

  • Magdalena Jachymek; 
  • Michał Tomasz Jachymek; 
  • Radosław Marek Kiedrowicz; 
  • Jarosław Kaźmierczak; 
  • Małgorzata Peregud-Pogorzelska

ABSTRACT

Background:

Recent advances in mobile sensor technology have led to increased popularity of wrist-worn fitness trackers. The possibility to use a smartwatch as a rehabilitation tool to monitor patients’ heart rate during exercise has won the attention of many researchers.

Objective:

The aim of the study was to evaluate the accuracy and precision of HR measurement performed by two wrist monitors: Fitbit Charge 4 (Fitbit) and Xiaomi Mi Band 5 (Xiaomi).

Methods:

31 healthy volunteers were asked to perform a stress test on a treadmill. During the test their heart rate was recorded simultaneously by both wristbands and ECG at 1minute intervals. The mean absolute error percentage (MAPE), Lin’s concordance correlation coefficient (LCCC) and Bland-Altman were calculated to compare precision and accuracy of heart rate measurements. The estimated validation criteria were MAPE < 10% and LCCC < .8

Results:

The overall MAPE of the Fitbit device was 10.19% (±11.79%) and the MAPE of Xiaomi was (6.89 % ± 9.75). LCCC of Fitbit HR measurements was .753 (95% CI:0.717-0.785) and of Xiaomi – .903 (0.886-0.917). In both devices the precision and accuracy were decreasing with the increasing exercise intensity. Age, sex, height, weight, BMI did not influence the accuracy of both devices.

Conclusions:

The accuracy of a wearable wrist-worn heart rate monitor varies and depends on the intensity of training. The decision concerning the application of such a device as a monitor during in-home rehabilitation should be taken with caution, as it may prove not reliable enough.


 Citation

Please cite as:

Jachymek M, Jachymek MT, Kiedrowicz RM, Kaźmierczak J, Peregud-Pogorzelska M

Validation of heart rate measurement of Fitbit Charge 4 and Xiaomi Mi Band 5

JMIR Preprints. 02/06/2021:30893

DOI: 10.2196/preprints.30893

URL: https://preprints.jmir.org/preprint/30893

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.