Accuracy, Repeatability, and Hypertension Classification of a Card-Sized, Cuff-Calibrated Cuffless Blood Pressure Monitor: Prospective Two-Stage Validation Study

  • Zhongyang Shi; 
  • Shuwen Yu; 
  • Rui Zhang; 
  • Rui Zhang; 
  • Xi Zhang; 
  • Rong Li; 
  • Jing Gao; 
  • Zhen Li; 
  • Huaqin Wang; 
  • Qing Zhao; 
  • Jing Li; 
  • Fengrui Yang; 
  • Rong Wang

ABSTRACT

Background:

Many cuffless blood pressure (BP) monitors need calibration with a cuff device. When test readings are compared with the calibration reading, or are taken only minutes after calibration, agreement may reflect the calibration value rather than independent measurement.

Objective:

This study aimed to evaluate the accuracy, short-term repeatability, and hypertension classification of a card-sized, cuff-calibrated cuffless BP monitor (Card-BP) that combines finger photoplethysmography with single-lead electrocardiography.

Methods:

In this prospective, single-center, 2-stage study in Jinan, China, Card-BP was calibrated for each participant with an upper-arm oscillometric monitor. In stage 2 (n=88 after exclusion of 69 of 157 measured participants for reference BP variability, as specified by ISO 81060-2), 2 observers took simultaneous mercury auscultatory readings alternating with 3 Card-BP and 3 oscillometric readings, to assess accuracy, repeatability, and performance relative to a single-reading comparator (the first oscillometric reading carried forward unchanged). In stage 1 (n=726), a Card-BP reading taken immediately after calibration was compared with the calibration reading. Classification of BP ≥140/90 mm Hg was examined in both stages.

Results:

In stage 2 (264 paired readings), Card-BP minus mercury differences were mean 0.53 (SD 7.72) mm Hg for systolic BP (SBP) and mean 1.62 (SD 6.24) mm Hg for diastolic BP (DBP), meeting the ISO 81060-2 numerical criteria. The single-reading comparator also met these criteria, with differences of mean −1.32 (SD 5.88) and mean 2.72 (SD 6.15) mm Hg, as did the oscillometric monitor. The SD of Card-BP differences minus that of the comparator was 1.84 (95% CI 0.65-3.15) mm Hg for SBP and 0.09 (95% CI −0.76 to 0.92) mm Hg for DBP. Card-BP repeatability coefficients were 11.4 mm Hg for SBP and 2.7 mm Hg for DBP (mercury: 9.4 to 9.6 and 6.8 to 7.2 mm Hg). In stage 1, Card-BP readings differed from the calibration reading by mean 0.01 (SD 7.90) mm Hg for SBP and mean −0.45 (SD 2.36) mm Hg for DBP. Among 291 stage 1 participants without diagnosed or treated hypertension, Card-BP classified hypertension against the calibration reading with a sensitivity of 89% (71/80) and a specificity of 93% (196/211). In stage 2, against mercury, sensitivity and specificity were 92% (22/24) and 98% (63/64) for Card-BP and 88% (21/24) and 100% (64/64) for the comparator.

Conclusions:

Within minutes of calibration, Card-BP met cuff-device accuracy criteria against mercury auscultation, but so did a single cuff reading carried forward unchanged, which agreed with the reference at least as closely. Card-BP diastolic readings changed little after calibration, and agreement measured against the calibration reading overstated accuracy. Validation of cuff-calibrated cuffless monitors should use reference readings independent of calibration, report a carry-forward comparator, and test the tracking of BP changes. Clinical Trial: Chinese Clinical Trial Registry ChiCTR2200066035; https://trialsearch.who.int/Trial2.aspx?TrialID=ChiCTR2200066035


 Citation

Please cite as:

Shi Z, Yu S, Zhang R, Zhang R, Zhang X, Li R, Gao J, Li Z, Wang H, Zhao Q, Li J, Yang F, Wang R

Accuracy, Repeatability, and Hypertension Classification of a Card-Sized, Cuff-Calibrated Cuffless Blood Pressure Monitor: Prospective Two-Stage Validation Study

JMIR Preprints. 20/11/2026:117285

DOI: 10.2196/preprints.117285

URL: https://preprints.jmir.org/preprint/117285

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.